Absolute Sensitivity of a Graphite Ionization Chamber
Phys. Rev. 72, 61 – Published 1 July, 1947
DOI: https://doi.org/10.1103/PhysRev.72.61
Abstract
A thick-walled graphite ionization chamber is suggested as a simple instrument for measuring the total intensity in the x-ray beam emitted by the betraton. The chamber would consist of two layers of carbon, each about 1/5 of a radiation length, or 4.5 cm thick, separated by a thin gas layer whose width is not critical. The transverse dimensions of the carbon layers should be sufficiently large to cover completely the x-ray beam as well as to catch the electrons produced at an angle in the carbon. The absolute sensitivity of the ionization chamber is computed to an accuracy of about 5 percent by comparison with the ionization which would be produced in an infinitely thick carbon block.
References (8)
- G. C. Laurence, Can. J. Research A15, 16 (1937)
- B. Rossi and K. Greisen, Rev. Mod. Phys. 13, 240 (1941)
- B. Rossi and K. Greisen, Rev. Mod. Phys. 13, 240 (1941) p. 247
- [3], p. 263
- L. H. Gray, Proc. Roy. Soc. A156, 578 (1936)
- [3], p. 251
- [3], p. 264
- [3], p. 256